The Intelligibility of the Reversed-Stethoscope Technique in Age-Related Hearing Loss

Raphaelle M. Koerber, PhD1, Linh Vaccarello, PhD3, Allan Ho, FRCS, MSc, MBBS2,3

1Michael G. DeGroote School of Medicine, McMaster University, Waterloo Regional Campus, Kitchener, ON
2Division of Otolaryngology Head and Neck Surgery, University of Alberta Faculty of Medicine and Dentistry, Edmonton, AB
3Edmonton Ear Clinic, Edmonton, AB

DOI: https://doi.org/10.5770/cgj.25.527


ABSTRACT

Background

This study evaluated the effectiveness of the reverse stethoscope technique in improving speech intelligibility. In this technique, a clinician places the earpieces of their stethoscope into the ears of a hearing-impaired patient and speaks into the chest piece.

Methods

The International Speech Test Signal was presented to four Littman® stethoscope models and a Pocketalker® personal voice amplifier using an Audioscan® hearing instrument test box. The acoustic outputs of the stethoscopes and voice amplifier were measured across the frequency spectrum of speech. The Speech Intelligibility Index of the resulting speech was calculated for natural speech and for each device in relation to 10 standardized hearing losses representing the population of older adults.

Results

For each of the 10 hearing losses, the speech signal emitted by the stethoscopes was quieter and yielded lower speech intelligibility scores than regular speech. In contrast, the voice amplifier provided mid- and high-frequency amplification and improved speech intelligibility scores for all but the mildest hearing losses.

Conclusions

The reverse stethoscope technique worsens the clarity of speech and should not be used with older, hearing-impaired patients. Instead, clinicians should use regular speech or, preferably, personal voice amplifiers.

Key words: hearing loss, health-care accessibility, stethoscope

INTRODUCTION

Among adults over 70, nearly two thirds have a hearing impairment.(1) In the geriatric patient population, this proportion is believed to be higher as individuals with hearing loss experience more chronic health conditions than their age-matched peers,(2) and are more frequently hospitalized.(3) While receiving care, hearing loss presents challenges. Pope et al.(4) found the speech comprehension of patients with hearing loss declined dramatically in the presence of hospital noise. Hearing aids can mitigate this problem, but almost four out of five octogenarians with significant hearing loss do not wear amplification devices.(5,6) Moreover, patients frequently lose their hearing aids during hospital admissions.(7) As a result, 75% of patients with hearing loss report sometimes or often misunderstanding their health-care providers.(8)

Personal voice amplifiers, such as the Pocketalker® (Williams AV, LLC., Eden Prairie, MD), are well recognized tools for improving communication. These battery-operated devices contain a microphone, amplifier, and headphones through which users listen. Both the volume and the pitch of the acoustic output can be modified to amplify to the degree and nature of the user’s hearing loss.(9) No research has quantified the speech intelligibility benefits of these devices. However, they have been found to improve quality of life and reduce depressive symptoms in older adults,(10,11) and to function reliably within care institutions.(12)

Within hospitals, a more readily available alternative to the personal voice amplifier is described in Samuel Shem’s 1978 novel ‘House of God’. Here, a senior resident, ‘The Fat Man’, teaches junior residents how to communicate with an elderly patient with a hearing loss:

‘With Anna you need the reverse stethoscope technique. Watch.’ The Fat Man took off his stethoscope, plugged the earpiece into Anna O.’s ears, and then, using the bell like a megaphone, shouted into it: ‘Cochlea come in, cochlea come in, do you read me…’(13)

While House of God was a satire, at times physicians use a version of this ‘reverse stethoscope’ strategy by speaking into the chest piece of an acoustic stethoscope with the earpieces placed in the patient’s ears. This method is described in a Medscape Perspective article as a strategy for communicating with elderly surgery patients.(14) It was recommended in the British Medical Journal’s Innovations section for a similar purpose.(15) The textbook “Emergency Care in the Streets” includes it as a tip for working with patients with hearing impairment.(16) A British Medical Journal practice pointer describing evidence-based strategies for communicating with patients with hearing loss did not include this technique,(17) a fact which three clinicians wrote in to comment on via the article’s “Rapid Responses”. Despite this technique’s popularity, the intelligibility of speech when delivered through acoustic stethoscopes has never been scientifically evaluated.

Stethoscopes are designed to transmit low-frequency cardiac, respiratory, and gastrointestinal sounds(18) Acoustic stethoscopes aid in the auscultation of these sounds in three ways. First, by pressing the chest piece against the human body, the device overcomes tissues’ acoustic impedance (i.e., resistance to sound wave transmission between tissue and air).(19) Second, by funneling sound waves into the tubing of the acoustic stethoscope, a greater proportion of the sound energy is directed towards the ear.(20,21) Finally, some—but not all—models provide resonance peaks of sufficient amplitude to amplify certain very low-pitched frequencies, below 200 Hz.(22,23,24) For sound waves with frequencies higher than 200 Hz, the intensity (i.e., loudness) of the delivered sound waves decreases steadily up until 1000 or 1500 Hz, at which point they can no longer be appreciated by human hearing. These outputs may not fully reflect the acoustics characteristics of the reverse stethoscope technique, as the stethoscope diaphragm vibrates in free air rather than pressed against a body surface.(25,26) Still, as most speech sounds exist between 500 and 8000 Hz,(27) this pattern of declining output calls the reverse stethoscope technique into question.

Stethoscopes’ lack of mid- and high-frequency output is particularly problematic for patients with age-related hearing loss, who experience a greater degree of loss in the high frequencies. For older adults, the high frequency sounds, including the /f/, /th/, and /s/ phonemes, are more difficult to discriminate than lower pitched vowel sounds and the sonorant consonants such as the /m/ and /r/ phonemes. This is visually demonstrated through the 10 ‘standard audiograms’ in Figures 1 and 2. These standard audiograms were developed by vector quantization of almost 30,000 audiograms from older adults with hearing loss.(28) In each of these standard hearing losses, thresholds worsen as the frequency of the test tone increases.

 


 

FIGURE 1 Flat and moderately sloping standard audiograms

 


 

FIGURE 2 Steeply sloping standard audiograms

Given the diversity of hearing losses, the Speech Intelligibility Index is one way to quantify the degree to which a hearing device provides appropriate amplification for a given loss. This index, based on calculations outlined by the American National Standards Institute,(29) estimates the percentage of daily speech sounds which a person with a given hearing loss can hear. It can be calculated to reflect the speech intelligibility of standard speech, as well as the intelligibility of amplified speech emitted by an assistive hearing device.

Existing literature suggests that stethoscopes attenuate, rather than amplify, the mid-to-high frequencies in which speech is found. This leads us to doubt the efficacy of the reverse stethoscope technique and ask the following questions:

  1. Does the Speech Intelligibility Index for 10 standard audiograms improve when the International Speech Test Signal is presented through four models of Littmann® stethoscopes (3M Littmann Stethoscopes, St. Paul, MN) in a test-box setup mimicking the “reverse stethoscope” technique?

  2. How do these scores compare to the Speech Intelligibility Index scores associated with a Pocketalker personal voice amplifier?

METHODS

Materials and Set Up

To evaluate the stethoscopes’ acoustic outputs, we used the Audioscan® Verifit2 test box (Audioscan, Dorchester, ON) and software. Best practices within the field of hearing-loss amplification recommend that hearing instruments and other assistive devices be evaluated relative to the user’s specific hearing loss using a test box and associated sound processing software.(18) One widely used test box system is the Audioscan Verifit2. This noise-insulated box contains three core components: a speaker emitting a standardized signal calibrated by a reference microphone, a coupler designed to hold the hearing instrument’s earpieces and mimic the acoustics of an ear, and a microphone positioned within this coupler.

In using the Audioscan Verifit2 to test the reversed stethoscopes, the adult-sided bell of each Littmann stethoscope was positioned within the test box facing the speaker at a distance of 2 cm. Due to their size, each consecutive stethoscope extended out of the test box with the metal component of the stethoscope tubing passing under the foam-sealed, test-box lid. The coupler likewise extended outside the test box via an extender cable. The earpieces of the stethoscope were then puttied to the 0.4 cc coupler, following standard practice for in-the-ear hearing aid verification (Audioscan Verifit, 2019). Foam was used to seal the lid, and the reference microphones within the test box were re-calibrated to this arrangement to account for any residual signal leakage.

In this configuration, test signals were presented to a Littmann Classic II SE stethoscope, a Littmann Cardiology IV stethoscope, and a Littmann Cardiology IV stethoscope with short-length tubing (56 cm vs. 69 cm). In addition, test signals were presented to a Littmann 3100 Electronic stethoscope in diaphragm (i.e., high frequency) mode, and with amplification turned up to level nine (i.e., the highest level of amplification provided by the device). This electronic stethoscope was included to reference to the speech intelligibility that amplified stethoscopes might offer. All stethoscopes’ outputs were measured by the microphones within the test box’s 0.4 cc couplers.

The acoustic output of the Pocketalker personal voice amplifier was evaluated using a comparable set-up. The hand-held unit was placed within the box, with the microphone facing the speaker at a distance of 2 cm. The device’s acoustic output was presented to the 0.4 cc coupler microphone via a mono earbud puttied to the coupler. The output of the stethoscopes and personal voice amplifier were interpreted by the Audioscan software to 1) provide spectral analyses of each devices’ outputs, and 2) calculate the Speech Intelligibility Index scores associated with each device relative to each of the 10 standard hearing losses.

Spectral Analyses

The output of the stethoscopes in dB HL was measured for frequencies between 125 and 13,000 Hz. We used Pink Noise as the input to evaluate the stethoscopes’ frequency response, following the methods of Weiss et al.(18,31) A 65 dB Pink Noise signal was presented via the test box speaker to the devices, captured by the coupler microphone, and presented using the ‘Multicurve’ function of the Audioscan software.

Speech Intelligibility Index Measurement

The Audioscan software was configured to measure the Speech Intelligibility Index for each of 10 standard audiograms reflecting the population of hearing-impaired older adults (see Figures 1 and 2). The Speech Intelligibility Index was calculated both at baseline, to reflect natural speech, and in response to speech presented through the reversed stethoscope. Software settings were chosen to improve accuracy, as outlined in Table 1.

TABLE 1 Audioscan speechmap configuration

 

Procedure

The International Speech Test Signal was delivered through the test box speakers at 65 dB HL, reflecting the intensity of average conversational speech (Audioscan Verifit, 2019), and filtered through each of the stethoscopes and the Pocketalker one at a time. The software interpreted the output for each device relative to each of the 10 standard losses outlined in Figures 1 and 2.(28) Each standard loss’ baseline “unaided” Speech Intelligibility Index was also calculated by the Audioscan software. This unaided Speech Intelligibility Index reflected the percentage of speech a person with a standard audiogram could hear without any assistive hearing device.

Data Analyses

The resulting spectral analyses and Speech Intelligibility Index scores were compared using descriptive statistics, as is standard in stethoscope acoustics research.(18,25,30,31)

RESULTS

Spectral Analyses

Spectral analyses of the Pocketalker and four stethoscopes demonstrated that the Pocketalker provided amplification across all frequencies involved in speech, while the acoustic stethoscopes only attenuated these signals (see Figure 3). The Pocketalker, at maximum volume, amplified the 65 dB input speech to an output ranging from 72 dB to 100 dB across the frequency spectrum. When the high-frequency setting was activated via the Pocketalker tone dial, the device provided 9 dB of additional amplification in the high frequencies and roughly 10 dB less amplification in the lower frequencies. Conversely, the only stethoscope which provided any amplification of the speech signal was the Littmann 3100 Electronic Stethoscope. At its loudest setting, this device provided an output signal 5 dB louder than the 65 dB input at both 250 and 400 Hz. At other frequencies, the electronic stethoscope’s output was attenuated (i.e., quieter) relative to the incoming speech.

 


 

FIGURE 3 Spectral analyses of stethoscopes’ and the personal voice amplifier’s output in response to 65 dB Pink Noise

Speech Intelligibility

Speech signals delivered through the acoustic stethoscopes yielded poorer Speech Intelligibility Index scores than regular speech (see Figures 4 and 5). The Littmann 3100 Electronic stethoscope similarly degraded the speech intelligibility in all standard hearing losses, apart from hearing losses N4 and N5 in which it provided a near negligible 4% and 1% improvement, respectively. In contrast, the output of standardized speech presented through the Pocketalker yielded improved speech intelligibility relative to unaided standard speech in all but the mildest standardized hearing losses, as demonstrated in Figures 4 and 5. The benefit provided by the Pocketalker ranged from a 5% loss in Speech Intelligibility Index scores for hearing loss N1, the mildest hearing loss, to a 59% gain in the Speech Intelligibility Index score for N4, a moderately severe hearing loss.

 


 

FIGURE 4 Speech Intelligibility Index scores for flat and moderately sloping standard audiograms, ranging from mild loss (N1) to profound loss (N7) in response to 65 dB standard speech input into Littman stethoscopes and Pocketalker personal voice amplifier

 


 

FIGURE 5 Speech Intelligibility Index scores for severely sloping standard audiograms, ranging from mild (S1) to severe (S3), in response to 65 dB standard speech input into Littman stethoscopes and the Pocketalker personal voice amplifier

DISCUSSION

This study sought to determine whether the Speech Intelligibility Index for 10 standard audiograms improved when the International Speech Test Signal was presented through Littmann stethoscopes in a test box set-up mimicking the “reverse stethoscope” technique. We also sought to evaluate how a personal voice amplifier would impact speech intelligibility in a comparable environment. We found that the reverse stethoscope reduced, rather than improved, Speech Intelligibility Index scores. In contrast, a Pocketalker personal amplifier yielded improved speech intelligibility relative to standard speech in all but the mildest hearing losses.

These findings are in keeping with our spectral analyses of the stethoscopes. While the Pocketalker personal amplification device provided up to 38 dB of amplification across the spectrum of frequencies, the acoustic stethoscopes attenuated the speech signal in all frequencies, and the electronic stethoscope only provided modest amplification in the lowest frequencies. The spectral output of our devices is consistent with the existing literature, which has identified resonance peaks in stethoscopes below 500 Hz, followed by a progressive attenuation of output, reaching 20 dB of attenuation of all input signals beyond 1,000 Hz.(22,23,24,32) Together, these findings support the assertion that stethoscopes are designed to transmit the low frequencies sound waves which contain internal body sounds, rather than the mid- to high-frequency sound waves which contain speech sounds.

While the reversed stethoscope technique is not supported, this study demonstrates the benefits to speech-intelligibility provided by personal voice amplifiers when communicating with patients with age-induced hearing loss. It clarifies why previous research has found that these personal voice amplifiers improve communication and psychosocial outcomes in older adults.(10,11,12) Patients only require basic training in the use of the Pocketalker, and Gilligan and Weinstein(33) have published instructions explaining the devices’ use at a grade five reading level.

If a personal voice amplifier is not available, other strategies exist which improve communication in the presence of hearing loss. To improve speech comprehension, health-care providers should reduce ambient noise,(34) face the patient when speaking,(35) articulate consonants carefully,(36) and identify patients with hearing loss using signage within charts or at the bedside.(37) Transcription and amplification technologies in the form of iPhones or Android apps also have potential.(38)

This study demonstrates that the reverse stethoscope technique worsens rather than improves Speech Intelligibility Index scores. However, our research was performed in a lab setting using standardized hearing losses, rather than patients in an authentic hospital setting. As a result of this limitation, it is possible that we failed to capture certain aspects of the practice. For example, stethoscope earpieces may block distracting ambient hospital noise. Likewise, while using this technique, clinicians are forced to face the patient, which can facilitate speech reading.(35)

Alternatively, this “reverse stethoscope” technique may have persisted simply by encouraging patients to ‘nod along’ instead of expressing their concerns to the physician. Patients with hearing loss are known to cope by withdrawing rather than drawing attention to their disability. This stems from a desire both to avoid being associated with the stigmatized disability, and to avoid inconveniencing conversation partners.(39,40,41) Future research might explore this and other reasons for which this acoustically ineffective technique continues to be used.

CONCLUSION

In conclusion, three quarters of adults with hearing loss report sometimes or often misunderstanding their health-care providers.(8) The reverse stethoscope technique has been described as a ‘megaphone’ for amplifying a physician’s voice for older patients.(13) Unfortunately, our findings confirm that stethoscopes do not amplify or even effectively transmit the frequencies important to speech. Instead, this technique decreases Speech Intelligibility Index scores. Conversely, the Pocketalker personal voice amplifier meaningfully improved Speech Intelligibility Index scores for all but the mildest hearing losses. Older adults with hearing loss have a moral and legal right to accessible health care.(42,43) Based on this study’s findings, accessible communication for older adults with hearing loss is best pursued through personal voice amplifiers.

ACKNOWLEDGEMENTS

Not applicable.

CONFLICT OF INTEREST DISCLOSURES

We have read and understood the Canadian Geriatrics Journal’s policy on conflicts of interest disclosure and declare that there are no conflicts of interest.

FUNDING

Funding provided by the Edmonton Ear Clinic.

REFERENCES

1 Lin FR, Thorpe R, Gordon-Salant S, Ferrucci L. Hearing loss prevalence and risk factors among older adults in the United States. J Gerontol A Biol Sci Med Sci. [Internet]. 2011 May 1 [cited 2021 Feb 16]; 66(5):582–90. Available from: https://doi.org/10.1093/gerona/glr002
cross-ref  

2 Pandhi N, Schumacher JR, Barnett S, Smith MA. Hearing loss and older adults’ perceptions of access to care. J Community Health [Internet]. 2011 Oct 1 [cited 2021 Feb 16]; 36(5):748–55. Available from: https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3197225/
cross-ref  

3 Genther DJ, Frick KD, Chen D, Betz J, Lin FR. Association of hearing loss with hospitalization and burden of disease in older adults. JAMA [Internet]. 2013 Jun 12 [cited 2021 Feb 16]; 309(22):2322–24. Available from: https://jamanetwork.com/journals/jama/fullarticle/1696091
cross-ref  

4 Pope DS, Gallun FJ, Kampel S. Effect of hospital noise on patients’ ability to hear, understand, and recall speech. Res Nurs Health [Internet]. 2013 Jun [cited 2021 Feb 16]; 36(3):228–41. Available from: https://doi.org/10.1002/nur.21540
cross-ref  

5 Chien W, Lin FR. Prevalence of hearing aid use among older adults in the United States. Arch Intern Med [Internet]. 2012 Feb 13 [cited 2021 Feb 16]; 172(3):292–93. Available from: doi:10.1001/archinternmed.2011.1408
cross-ref  

6 Nieman CL, Marrone N, Szanton SL, Thorpe Jr RJ, Lin FR. Racial/ethnic and socioeconomic disparities in hearing health care among older Americans. J Aging Health [Internet]. 2016 Feb [cited 2021 Feb 16]; 28(1):68–94. Available from: https://doi.org/10.1177/0898264315585505
cross-ref  pmc  

7 Oliver D. Admission should allow for patient aids. BMJ [Internet]. 2017 Sep 19 [cited 2021 Feb 16]; 358. Available from: https://www.bmj.com/content/358/bmj.j3953.full
cross-ref  

8 Stevens MN, Dubno JR, Wallhagen MI, Tucci DL. Communication and healthcare: self-reports of people with hearing loss in primary care settings. Clin Gerontol [Internet]. 2019 Oct 20 [cited 2021 Feb 16]; 42(5):485–94. Available from: https://doi.org/10.1080/07317115.2018.1453908
cross-ref  

9 William Sound. Specification Data Pocketalker® Ultra Personal Amplifier [Internet]. Eden Prairie, MN: William Sound; 2019 [cited 2021 Feb 16]. Available from: https://www.williamssound.com/resources/products/web/pkt/pkt_d1_spec_en.pdf

10 Mamo SK, Nirmalasari O, Nieman CL, et al. Hearing care intervention for persons with dementia: a pilot study. Am J Geriatr Psychiatry [Internet]. 2017 Jan 1 [cited 2021 Feb 16]; 25(1):91–101. Available from: https://doi.org/10.1016/j.jagp.2016.08.019
cross-ref  

11 Nieman CL, Marrone N, Mamo SK, et al. The Baltimore HEARS Pilot Study: an affordable, accessible, community-delivered hearing care intervention. Gerontologist [Internet]. 2017 Nov 10 [cited 2021 Feb 16]; 57(6):1173–86. Available from: https://doi.org/10.1093/geront/gnw153

12 Lewsen BJ, Cashman M. Hearing aids and assistive listening devices in long-term care. Can J Speech Lang Pathol Audiol [Internet]. 1997 [cited 2021 Feb 16]; 21(3):149–52. Available from: https://cjslpa.ca/files/1997_JSLPA_Vol_21/No_03_145-228/Lewsen_Cashman_JSLPA_1997.pdf

13 Shem S. The House of God. New York: Berkley Books; 2010.

14 Lowenfels AB. Improving the outcomes of major abdominal surgery in elderly patients. Medscape Perspectives [Internet]. 2005 Nov 22 [cited 2021 Feb 16]. Available from: https://www.medscape.com/viewarticle/517308

15 Vadodaria B. An unusual use of a stethoscope. BMJ [Internet]. 1998 May 2 [cited 2021 Feb 16]; 316(7141):1382. Available from: https://www.bmj.com/content/316/7141/1382
cross-ref  

16 Caroline NL. Emergency Care in the Streets. Sudbury, MA: Jones & Bartlett Learning; 2008 [cited 2021 Feb 16].

17 Middleton A, Niruban A, Girling G, Myint PK. Communicating in a healthcare setting with people who have hearing loss. BMJ [Internet]. 2010 Sep 29 [cited 2021 Feb 16]; 341:c4672. Available from: https://doi.org/10.1136/bmj.c4672

18 Weiss D, Erie C, Butera J III, et al. An in vitro acoustic analysis and comparison of popular stethoscopes. Med Devices (Auckl) [Internet]. 2019 [cited 2021 Feb 16]; 12:41–52. Available from: https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6339642/

19 Murphy A, Morgan MA. Acoustic impedance [Internet]. Radiopaedia.org. 2020, March 1 [cited 2021 Feb 16]. Available from https://radiopaedia.org/articles/acoustic-impedance?lang=us

20 Groom D. Comparative efficiency of stethoscopes. Am Heart J [Internet]. 1964 Aug 1 [cited 2021 Feb 16]; 68(2):220–26. Available from: https://www.sciencedirect.com/science/article/abs/pii/0002870364900432
cross-ref  

21 Rappaport MB, Sprague HB. Physiologic and physical laws that govern auscultation, and their clinical application: the acoustic stethoscope and the electrical amplifying stethoscope and stethograph. Am Heart J [Internet]. 1941 Mar 1 [cited 2021 Feb 16]; 21(3):257–318. Available from: https://doi.org/10.1016/S0002-8703(41)90904-3
cross-ref  

22 Ertel PY, Lawrence M, Brown RK, Stern AM. Stethoscope acoustics: I. The doctor and his stethoscope. Circulation [Internet]. 1966 Nov [cited 2021 Feb 16]; 34(5):889–98. Available from: https://www.ahajournals.org/doi/pdf/10.1161/01.CIR.34.5.889
cross-ref  

23 Kindig JR, Beeson TP, Campbell RW, Andries F, Tavel ME. Acoustical performance of the stethoscope: a comparative analysis. Am Heart J [Internet]. 1982 Aug 1 [cited 2021 Feb 16]; 104(2):269–75. Available from: https://doi.org/10.1016/0002-8703(82)90203-4
cross-ref  

24 Abella M, Formolo J, Penney DG. Comparison of the acoustic properties of six popular stethoscopes. J Acoust Soc Am [Internet]. 1992 Apr [cited 2021 Feb 16]; 91(4):2224–28. https://doi.org/10.1121/1.403655
cross-ref  

25 Nowak LJ, Nowak KM. Sound differences between electronic and acoustic stethoscopes. Biomed Eng Online [Internet]. 2018 Dec [cited 2021 Feb 16]; 17(1). Available from: https://doi.org/10.1186/s12938-018-0540-2

26 Leng S, San Tan R, Chai KT, Wang C, Ghista D, Zhong L. The electronic stethoscope. Biomed Eng Online [Internet]. 2015 Dec [cited 2021 Feb 16]; 14(1):1–37. Available from: https://doi.org/10.1186/s12938-015-0056-y

27 Fant G. Speech Acoustics and Phonetics: Selected Writings. Dordrecht: Kluwer Academic Publishers; 2004.

28 Bisgaard N, Vlaming MS, Dahlquist M. Standard audiograms for the IEC 60118-15 measurement procedure. Trends Amplif [Internet]. 2010 Jun [cited 2021 Feb 16]; 14(2):113–20. Available from: https://doi.org/10.1177/1084713810379609
cross-ref  

29 Acoustical Society of America. Methods for Calculation of the Speech Intelligibility Index (Standard No S3.-1997(R2017)) [Internet]. New York City: American National Standards Association; 1997 [cited 2021 Feb 16]. Available from: https://webstore.ansi.org/standards/asa/ansiasas31997r2017

30 Jorgensen LE. Verification and validation of hearing aids: opportunity not an obstacle. J Otol [Internet]. 2016 Jun 1 [cited 2021 Feb 16]; 11(2):57–62. Available from: https://doi.org/10.1016/j.joto.2016.05.001
cross-ref  

31 Pavlosky A, Glauche J, Chambers S, Al-Alawi M, Yanev K, Loubani T. Validation of an effective, low cost, Free/open access 3D-printed stethoscope. PLoS One [Internet]. 2018 Mar 14 [cited 2021 Feb 16]; 13(3):e0193087. Available from: https://doi.org/10.1371/journal.pone.0193087
cross-ref  

32 Ertel PY, Lawrence M, Brown RK, Stern AM. Stethoscope acoustics: II. Transmission and filtration patterns. Circulation [Internet]. 1966 Nov [cited 2021 Feb 16]; 34(5):899–909. Available from: https://www.ahajournals.org/doi/abs/10.1161/01.CIR.34.5.899
cross-ref  

33 Gilligan J, Weinstein BE. Incorporating health literacy into your hearing care practice. Hearing Rev [Internet]. 2016 Sept 15 [cited 2021 Feb 16]. Available from: https://www.hearingreview.com/practice-building/practice-management/incorporating-health-literacy-hearing-care-practice

34 Killion MC, Niquette PA, Gudmundsen GI, Revit LJ, Banerjee S. Development of a quick speech-in-noise test for measuring signal-to-noise ratio loss in normal-hearing and hearing-impaired listeners. J Acoust Soc Am [Internet]. 2004 Oct [cited 2021 Feb 16]; 116(4):2395–405. Available from: https://doi.org/10.1121/1.1784440
cross-ref  

35 Middelweerd MJ, Plomp R. The effect of speechreading on the speech-reception threshold of sentences in noise. J Acoust Soc Am [Internet]. 1987 Dec [cited 2021 Feb 16]; 82(6):2145–7. Available from: https://doi.org/10.1121/1.395659
cross-ref  

36 Caissie R, Tranquilla M. Enhancing conversational fluency: training conversation partners in the use of clear speech and other strategies. Semin Hear [Internet]. 2010 May [cited 2021 Feb 16]; 31(2): 95–103. Available from: https://www.thieme-connect.com/products/ejournals/abstract/10.1055/s-0030-1252101
cross-ref  

37 Fook L, Morgan R, Sharma P, Adekoke A, Turnbull CJ. The impact of hearing on communication. Postgrad Med J [Internet]. 2000 Feb 1 [cited 2021 Feb 16]; 76(892):92–95. Available from: https://pmj.bmj.com/content/76/892/92.short
cross-ref  

38 Hearing Link. Useful apps for hearing loss [Internet]. 2019 [cited 2021 Feb 16]. Available from: https://www.hearinglink.org/living/loops-equipment/useful-apps-for-hearing-loss/

39 Hallam RS, Ashton P, Sherbourne K, Gailey L, Corney R. Coping, conversation tactics and marital interaction in persons with acquired profound hearing loss (APHL): correlates of distress. Audiol Med [Internet]. 2007 Jan 1 [cited 2021 Feb 16]; 5(2):103–11. Available from: https://doi.org/10.1080/16513860701223060
cross-ref  

40 Hetu R, Riverin L, Getty L, Lalande NM, St-Cyr C. The reluctance to acknowledge hearing difficulties among hearing-impaired workers. Br J Audiol [Internet]. 1990 Jan 1 [cited 2021 Feb 16]; 24(4):265–76. Available from: https://doi.org/10.3109/03005369009076565
cross-ref  

41 Southall K, Gagné JP, Jennings MB. Stigma: a negative and a positive influence on help-seeking for adults with acquired hearing loss. Int J Audiol [Internet]. 2010 Nov 1 [cited 2021 Feb 16]; 49(11):804–14. Available from: https://doi.org/10.3109/14992027.2010.498447
cross-ref  

42 AMERICANS WITH DISABILITIES ACT (ADA) OF 1990, Pub. L. 101–336, 104 Stat. 327 (July 26, 1990). Available from: https://www.govinfo.gov/content/pkg/STATUTE-104/pdf/STATUTE-104-Pg327.pdf

43 ACCESSIBLE CANADA ACT OF 2019, S.C. 2019, C. 10 (June 21, 2021). Available from: https://www.canlii.org/en/ca/laws/astat/sc-2019-c-10/latest/sc-2019-c-10.html#:~:text=SUMMARY,persons%20with%20disabilities%2C%20in%20society


Correspondence to: Raphaelle Koerber, md, phd, Michael G. DeGroote School of Medicine, McMaster University, Waterloo Regional Campus, 10-B Victoria Street South, 3rd Floor, Kitchener, ON N2G 1C3, E-mail: raphaelle.koerber@medportal.ca